Cooling Incubators

Refrigerated incubator IC 150 R

Refrigerated incubator IC 150 R

Features

Working range: 0°C ... +60°C (IC 150-R)

Working range: 5°C* ... +60°C (IC 50-P)
*at least 20°C below ambient temperature

7-inch colour touchscreen display

Data logging and downloading via USB key

Alarm and operation logging

Temperature ramps and graphic display

8 storable programmes with 8 steps each

Single-step basic programme

Multi-step programme up to 64 steps

Other Features

ArgoLab forced ventilation refrigerated incubator.

Steel outer door with large glass window and interior light for sample observation (IC 150-R).

Double door with a glass internal door for IC 50-P.

Simultaneous display of all operating parameters.

Operator-silenced visual and audible alarms.

Digital timer 1 min - 99 h 59 min and continuous operation.

Delayed start' function of the heating cycle.

Safe temperature' function: working temperature limitation for sample protection.

Working data logging of up to 200.000 values and download to USB stick as CSV or PDF file.

Alarm and operation logging.

Temperature ramp function for specialised applications.

Graphical display of temperature trends.

Multi-user management with 3 authorisation levels.

Safety class 3.1 (DIN 12880): maximum temperature controller with fluid expansion.

Lateral passage hole 25 mm internal diameter as standard for insertion of one or more external sensors for internal temperature control and/or certification.

Stainless steel chamber with rounded corners, fully removable anti-tip supports and shelves for easy cleaning.

Power supply 220 ± 10% V - 50 Hz.

Large backlit LCD display, 7 storable programmes with 10 steps each (temperature, time, ventilation speed), a multi-step programme with up to 64 and an additional basic single-step programme (temperature, time, speed and ventilation).

Included Accessories

Steel grid shelves (2 for IC 50-P, 3 for IC 150-R), calibration report at 25°C performed with Accredia-referenced sample instrument and USB stick for data recording and downloading.
Product Code
41201512
Category
Cooling Incubators
Capacity
150
Maximum shelf load
10
Security class
3.1
Through-hole diameter
25
Dimension
650 x 620 x 1350
Frequency
50 - 60
Warranty
2
Working method
Forced ventilation
Weight
100
Absorbed power
1100
Programs
8 programs with up to 8 steps each, 1 single-step program, and 1 multi-step program with up to 64 steps
Coolant quantity
280
Coolant type
R1234ZE
Temperature resolution
0,1
Data transfer
Yes, via USB stick
Heating time (25°C)
40
Timer
99 : 59
Ventilation
3 speeds (50 %, 75 %, 100 %)
Temperature variation at 25 °C
±0,1
Useful volume
150
Power supply
230
Internal dimensions
500 x 360 x 800
Overtemperature security
Electronic and electromechanical
Working temperature
0...60
Display
Touchscreen
Regulations
2006/95/EC -2004/108/EC - EN61326-1:2006 - EN61010-1:2010 - EN61010-2-010:2003
Temperature homogeneity at 25 °C
±0,5
Packaging dimensions (W x D x H)
750 x 760 x 1600
Temperature ramps
Yes
Multi user
Yes
Alarm History
Yes
Operations history
Yes
Internal data logger
Yes
Maximum number of data
200000
Graph
Yes
Multi-step program
Yes
Maximum number of shelves
9