Forced Ovens

TCF 200 PLUS forced ventilation oven

TCF 200 PLUS forced ventilation oven

Features

Working range: +10°C Ta ... +300°C

LCD display

Data download and recording via USB key

8 storable programmes with 8 steps each

Single-step basic programme

Multi-step programme with up to 64 steps

Other Features

ArgoLab forced ventilation oven.

Simultaneous display of all operating parameters.

Visual and acoustic alarms that can be muted by the operator.

Digital timer 1 min - 99 h 59 min and continuous operation.

Delayed start’ function of the heating cycle.

Safe temperature' function: working temperature limitation for sample protection.

Fan adjustable to 3 levels: High = 100% - Medium = 75% - Low = 50%.

Minimal heating times and low heating power for low energy consumption in every application.

Working data logging of up to 200.000 values and download to USB stick.

Safety class 3.1 (DIN 12880): fluid expansion maximum temperature controller.

Through-hole Ø 5 mm as standard (upper side) for insertion of an external sensor for internal temperature control and/or certification.

Stainless steel chamber with rounded corners, fully removable anti-tip supports and shelves for easy cleaning.

Door switch for easy sample loading and unloading.

Power supply 220 ± 10% V - 50 / 60 Hz.

Included Accessories

steel grid shelves (n°2 shelves for models up to 50 liters, n°3 shelves for larger models) calibration report at 150°C performed with Accredia-referenced sample instrument, USB stick for data recording and downloading.
Product Code
41200222
Category
Forced Ovens
Series
PLUS
Capacity
200
Maximum shelf load
20
Security class
3.1
Through-hole diameter
5
Dimension
945 x 825 x 870
Frequency
50 - 60
Warranty
2
Working method
Forced ventilation
Weight
103
Absorbed power
2400
Programs
8 programs with up to 8 steps each, 1 single-step program, and 1 multi-step program with up to 64 steps
Temperature resolution
0.1
Data transfer
Yes, via USB stick
Heating time (150°C)
30
Timer
99 : 59
Ventilation
3 speeds (50 %, 75 %, 100 %)
Temperature variation at 150 °C
± 0,4
Useful volume
200
Power supply
230
Internal dimensions
645 x 495 x 650
Overtemperature security
Electronic and electromechanical
Working temperature
+10 Ta … 300
Display
LCD
Regulations
2006/95/EC -2004/108/EC - EN61326-1:2006 - EN61010-1:2010 - EN61010-2-010:2003
Temperature homogeneity at 150 °C
±3
Temperature ramps
No
Multi user
No
Alarm History
No
Operations history
No
Internal data logger
Yes
Maximum number of data
2167
Graph
No
Door switch
Yes
Multi-step program
Yes
Maximum number of shelves
9